| Model Number |
MF-410375 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
06/14/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture baker grade iii.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: weight to the spec, opening, crease fold and wear abrasion.A microscopic analysis was performed which identify a sharp edge opening assessed as unidentified tear opening.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a opening sharp in the anterior side assessed as unidentified (tear) opening.
|
| |
|
Event or Problem Description
|
|
The device has been explanted and replaced.
|
| |
|