Model Number 60-01-00 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/27/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a centrifugal pump system with tubing clamp gave two error codes associated to difference between the set value and actual value of pump speed and to inconsistent/no impulse from the hall sensor during procedure.There was no report of patient injury.
|
|
Manufacturer Narrative
|
H.10: based on the currently available elements, considering the investigation results of similar complaints, the most likely root causes of the reported errors are associated to issues with motor control board and processor board.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See initial report.
|
|