• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRIO ST; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRIO ST; SURGICAL MESH Back to Search Results
Catalog Number 5950030
Medical Device Problem Codes Defective Device (2588); Patient-Device Incompatibility (2682); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Seroma (2069); Insufficient Information (4580)
Date of Event 06/02/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Should additional information be provided, a supplemental emdr will be submitted.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol ventrio st on (b)(6) 2020.As reported, the patient is making a claim for an adverse patient outcome against the ventrio st.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient experienced emotional distress and the device was defective.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol ventrio st on (b)(6) 2020.As reported, the patient is making a claim for an adverse patient outcome against the ventrio st.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient experienced emotional distress and the device was defective.Addendum per additional information provided: (b)(6) 2020 - patient was diagnosed with ventral incisional hernia, umbilical hernia thereby underwent open repair with the implant of the bard ventrio st mesh.Per operative notes, ¿a bard ventrio st mesh was placed and sutured.¿ (b)(6) 2020 - patient was diagnosed with incisional seroma, wound thereby underwent open repair with evacuation of an incisional seroma, wound washout.Per operative notes, ¿there was no recurrent hernia sac.The wound was irrigated.Wound vac was placed with good seal.¿.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Addendum: h11: this supplemental emdr is submitted to document additional information provided and to correct the implant date.Root cause is inconclusive, no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Updated fields: a2, a4, b3, b4, b5, b7, d4 (product catalog no., udi no.), g3, g6, h2, h6, h10, h11.Corrected field: d.6a (date of implant).Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRIO ST
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
4018258495
MDR Report Key12518071
Report Number1213643-2021-07640
Device Sequence Number19853718
Product Code FTL
UDI-Device Identifier00801741031526
UDI-Public(01)00801741031526
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K101920
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number5950030
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/20/2021
Supplement Date Received by Manufacturer07/08/2025
Initial Report FDA Received Date09/23/2021
Supplement Report FDA Received Date07/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
Patient SexFemale
Patient Weight124 KG
-
-