MAUDE Adverse Event Report: EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2N1333
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event occurred during the unspecified date of (b)(6) 2021.(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that at least four (4) non-dehp standard bore catheter extension sets were not irrigating and were not able to flush.This was identified during an unspecified process step.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
H10: only three (3) samples were received for evaluation.The other sample was not received and therefore, could not be evaluated.Visual inspection on the received samples did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed which included ensuring the slide clamp was open, priming, purging air, attached to a male luer until secure, and opening and closing the slide clamp.All three returned samples failed clear passaged testing and pressure testing.The reported condition was verified.The cause of the condition was due to a manufacturing issue.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
MDR Report Key12519061
Report Number1416980-2021-05873
Device Sequence Number14511113
Product Code FPA
UDI-Device Identifier00085412475394
UDI-Public(01)00085412475394
Combination Product (Y/N)Y
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K132734
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number2N1333
Device Lot NumberUR21D21029
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2021
Initial Date Received by Manufacturer 08/31/2021
Supplement Date Received by Manufacturer10/19/2021
Initial Report FDA Received Date09/23/2021
Supplement Report FDA Received Date10/25/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
-
-