| Catalog Number |
2N1333 |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event occurred during the unspecified date of (b)(6) 2021.(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that at least four (4) non-dehp standard bore catheter extension sets were not irrigating and were not able to flush.This was identified during an unspecified process step.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: only three (3) samples were received for evaluation.The other sample was not received and therefore, could not be evaluated.Visual inspection on the received samples did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed which included ensuring the slide clamp was open, priming, purging air, attached to a male luer until secure, and opening and closing the slide clamp.All three returned samples failed clear passaged testing and pressure testing.The reported condition was verified.The cause of the condition was due to a manufacturing issue.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|