BOSTON SCIENTIFIC CORPORATION SEGURA HEMISPHERE; DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number M0063801100
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(6).(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a segura hemi basket was used in the ureter during a flexible ureteroscopy procedure performed on (b)(6) 2021.During the procedure, the distal end of the sheath detached inside the patient.The detached sheath was retrieved using a foreign body clamp.Procedure was completed with this device.There was no patient complications as a result of this event.
 
Event or Problem Description
It was that a segura hemi basket was used in the ureter during a flexible ureteroscopy procedure performed on (b)(6) 2021.During the procedure, the distal end of the sheath detached inside the patient.The detached sheath was retrieved using a foreign body clamp.Procedure was completed with this device.There was no patient complications as a result of this event.
 
Additional Manufacturer Narrative
Block e1: the initial reporter phone number is (b)(6).Block h6: device code a0414 captures the reportable event of sheath torn material.Block h10: the returned segura hemi basket was analyzed, and a visual evaluation noted that the basket was returned retracted.One part of the sheath was returned detached and torn.Functional inspection found the basket was able to open or close.No other issues was noted.Based on all available information, it is possible that interaction with the scope and the physician's technique can cause sheath torn and detached at the distal section.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEGURA HEMISPHERE
Common Device Name
DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key12519087
Report Number3005099803-2021-04992
Device Sequence Number14567827
Product Code FFL
UDI-Device Identifier08714729341925
UDI-Public08714729341925
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 01/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/05/2024
Device Model NumberM0063801100
Device Catalogue Number380-110
Device Lot Number0027089098
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/02/2021
Supplement Date Received by Manufacturer01/04/2022
Initial Report FDA Received Date09/23/2021
Supplement Report FDA Received Date01/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-