| Model Number |
LIPC |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter received a questionable lipc lipase colorimetric result for one patient sample with cobas 6000 c501 module serial number (b)(4).On (b)(6) 2021, the initial result was 12,116 u/l.On (b)(6) 2021, a second sample was tested with a result of 65 u/l.Both results were reported outside of the laboratory.The physician questioned the results as he thought the value decreased too fast after treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
Qc trace was noted to be within normal ranges.As the patient's result after treatment was considered to be correct, a potential interference is unlikely.No reaction kinetics, instrument alarm report, or other test results from the patient were available to allow for further analysis.A general reagent problem can be excluded.No product problem was found.
|
| |
|