Model Number 1217-31-054
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 09/02/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that there was dissociation of the liner.There was surgical delay and needed additional surgery.Doi: (b)(6) 2021, dor: (b)(6) 2021, unknown side.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : examination of the returned device and available x-rays and photographs finds sufficient evidence to confirm the liner from disassociation event.A conclusive determination of root cause not identified.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN 100 W/GRIPTION 54MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12519637
Report Number1818910-2021-20925
Device Sequence Number5122405
Product Code KWA
UDI-Device Identifier10603295010210
UDI-Public10603295010210
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K090998
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1217-31-054
Device Catalogue Number121731054
Device Lot Number8746075
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2021
Initial Date Received by Manufacturer 09/02/2021
Supplement Date Received by Manufacturer09/20/2021
10/14/2021
10/25/2021
Initial Report FDA Received Date09/23/2021
Supplement Report FDA Received Date09/28/2021
10/14/2021
10/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ACTIS COLLARED HIGH SIZE 7; ALTRX NEUT 36IDX54OD; DELTA CER HEAD 12/14 36MM +8.5; PINN 100 W/GRIPTION 54MM; ACTIS COLLARED HIGH SIZE 7; ALTRX NEUT 36IDX54OD; DELTA CER HEAD 12/14 36MM +8.5; PINN 100 W/GRIPTION 54MM
Outcome Attributed to Adverse Event Required Intervention;
-
-