| Catalog Number |
07026773190 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of a questionable false-positive anti-hav igm elecsys result from cobas e 801 analytical unit serial number (b)(4).The initial result was 1.78 coi (positive).The repeat result was 1.75 coi (positive).The repeat result after centrifugation with a high-speed centrifuge was 1.71 coi (positive).The result from a siemens analyzer was 0.10 coi (negative).The questionable results were not reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
No sample material was available, therefore no further investigation was possible.The investigation did not identify a product problem. the cause of the event could not be determined. based on calibration and qc results, a general reagent issue could be excluded.
|
| |
|