| Model Number |
306601 |
| Medical Device Problem Codes |
Break (1069); Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received indicating that a smiths medical peripheral intravenous catheters (pivc)|jelco safety protectiv plus catheter appeared short upon removal.No patient consequences were reported.No adverse effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: one unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.However, as the samples were not returned with blister packaging and lot number(s) were unknown, causal factor could not be confirmed as a manufacturing related issue.Dhr review could not be conducted since the lot number is unknown.
|
| |
|