| Model Number |
3003-0822S |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported a gamma 3 set screw strayed into the body due to being buried outside the nail and was difficult to remove, so the operation was completed with the screw still in the body.There were no reported adverse consequences to the patient.
|
| |
|
Event or Problem Description
|
|
It was reported a gamma 3 set screw strayed into the body due to being buried outside the nail and was difficult to remove, so the operation was completed with the screw still in the body.There were no reported adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
|
| |
|