Model Number CR2
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2021
Event Type  malfunction  
Manufacturer Narrative
The customer has been provided with a replacement lid.The device nor the lid was returned to physio-control for evaluation.An engineering investigation has determined that due to the design of the lcpr2 lid switch, absence of the lid magnet allows current to flow from the battery, even when the device is in standby mode.This will reduce the life of the battery.Therefore, the reported issue, the loss of lid/lid magnet, is associated with two hazardous situations: opening the lid does not turn on the device, and, higher than intended current draw while the device is in standby mode results in a prematurely depleted the battery.Replacement of the device lid so that the magnet is present in the device allows the lid switch to function as designed; turning on/off the device when the lid is opened and preventing current draw when the device lid is closed.The lpcr2 operating instructions has been updated to include additional troubleshooting tips to direct the customer to act when device behavior indicates the lid magnet may be missing.A new warning informs the customer of the risk associated with a missing magnet.
 
Event Description
The customer contacted physio-control to report a non-critical issue with their device.Upon further investigation, physio-control was made aware that the device was missing the magnet from the lid of the device.In this state the device will not turn on automatically when the lid is activated, which can cause a delay in defibrillation therapy if needed or may lead to use error resulting in a failure to deliver defibrillation.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) CR2 DEFIBRILLATOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key12522816
MDR Text Key273090559
Report Number0003015876-2021-01824
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873824108
UDI-Public00883873824108
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P170018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date 09/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberCR2
Device Catalogue Number99512-001264
Was Device Available for Evaluation? No
Date Manufacturer Received08/30/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/22/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number3015876-01/14/2021-001-C
Patient Sequence Number1
-
-