Model Number 3660 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 09/08/2021 |
Event Type
Injury
|
Event Description
|
It was reported that during a revision of leads, ipg was placed in surgery mode prior to the procedure.During the procedure company representative were unable to recover the connection.Troubleshooting was done during the procedure and exhausted all steps.As such a new ipg was opened and replaced with the old ipg, thereby addressing the issue.
|
|
Manufacturer Narrative
|
The report of inoperable ipg was confirmed.Analysis of the returned ipg found the inability to communicate between the clinicians programmer and the implantable pulse generator was due to the device entering the service application state.This issue can occur when the device is exposed to an external energy source outside the device handling and storage requirements.Once the device has entered into the service application state, the cp will be inhibited from programming the device into the therapy application state.The ipg logs confirmed that the date the device entered into service app was the day of the lead revision surgery (b)(6) 2021.As a result of this finding, actions have been taken to prevent reoccurrence.There is sufficient instruction in the clinician's manual regarding use of electrosurgery devices during surgical procedures.
|
|
Manufacturer Narrative
|
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
|
|
Search Alerts/Recalls
|