ONKOS SURGICAL ELEOS; TIBIAL BASEPLATE, SIZE 2, 65MM M/L Back to Search Results
Model Number 25002202E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 08/27/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device will not be returned for investigation.The investigation is in progress.When the investigation is complete, a supplemental report will be submitted accordingly.Multiple mdrs were reported for this event: #3013450937-2021-00209, #3013450937-2021-00210, #3013450937-2021-00211, #3013450937-2021-00213, #3013450937-2021-00214, #3013450937-2021-00215, #3013450937-2021-00216, #3013450937-2021-00217.
 
Event or Problem Description
It was reported that the patient underwent a revision surgery on (b)(6) 2021 due to an alleged infection.The following part were removed and revised: distal femur axial pin, tibial hinge component, tibial poly spacer.The following parts remain implanted from the original eleos surgery which took place on (b)(6) 2021: tibial baseplate, distal femur, stem extension, collar, collar locking ring, and cemented stem.The surgeon removed implants with poly components and performed incision and drainage.No further information was made available.
 
Additional Manufacturer Narrative
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the alleged infection could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that could have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.
 
Event or Problem Description
It was reported that the patient underwent a revision surgery on (b)(6) 2021 due to an alleged infection.The following part were removed and revised: distal femur axial pin, tibial hinge component, tibial poly spacer.The following parts remain implanted from the original eleos surgery which took place on (b)(6) 2021: tibial baseplate, distal femur, stem extension, collar, collar locking ring, and cemented stem.The surgeon removed implants with poly components and performed incision and drainage.No further information was made available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
TIBIAL BASEPLATE, SIZE 2, 65MM M/L
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey rd
parsippany NJ 07054
Manufacturer (Section G)
MICROPORT ORTHOPEDICS
5677 airline road
arlington TN 38002
Manufacturer Contact
sara dailey
77 east halsey road
parsippany, NJ 07054
MDR Report Key12523113
Report Number3013450937-2021-00212
Device Sequence Number12805595
Product Code KRO
UDI-Device IdentifierB27825002202E0
UDI-Public+B27825002202E0
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K161520
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number25002202E
Device Catalogue Number25002202E
Device Lot Number1792981
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2021
Supplement Date Received by Manufacturer11/22/2021
Initial Report FDA Received Date09/23/2021
Supplement Report FDA Received Date11/22/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/13/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
P/N 25000007E, ELEOS DISTAL FEMUR; P/N 25001212E, ELEOS TIBIAL POLY SPACER; P/N 25002111E, ELEOS DISTAL FEMUR AXIAL PIN; P/N HC-11120-03M, ELEOS CEMENTED STEM; P/N HR-30000-03M, ELEOS COLLAR LOCKING RING; P/N KSC01530E, ELEOS STEM EXTENSION, 15X30MM; P/N SB-3640V-03M, ELEOS COLLAR, 36/40 OVAL, SOLID; P/N THSMWRS01M, ELEOS TIBIAL HINGE COMPONENT
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-