|
|
| Model Number |
KSC01530E |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
08/27/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device will not be returned for investigation.The investigation is in progress.When the investigation is complete, a supplemental report will be submitted accordingly.Multiple mdrs were reported for this event: #3013450937-2021-00209, #3013450937-2021-00210, #3013450937-2021-00211, #3013450937-2021-00212, #3013450937-2021-00213, #3013450937-2021-00215, #3013450937-2021-00216, #3013450937-2021-00217.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a revision surgery on (b)(6) 2021 due to an alleged infection.The following part were removed and revised: distal femur axial pin, tibial hinge component, tibial poly spacer.The following parts remain implanted from the original eleos surgery which took place on (b)(6) 2021: tibial baseplate, distal femur, stem extension, collar, collar locking ring, and cemented stem.The surgeon removed implants with poly components and performed incision and drainage.No further information was made available.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a revision surgery on (b)(6) 2021 due to an alleged infection.The following part were removed and revised: distal femur axial pin, tibial hinge component, tibial poly spacer.The following parts remain implanted from the original eleos surgery which took place on (b)(6) 2021: tibial baseplate, distal femur, stem extension, collar, collar locking ring, and cemented stem.The surgeon removed implants with poly comopents and performed incision and drainage.No further information was made available.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the alleged infection could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that could have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.The following sections were updated: h3: device evaluated by manufacturer updated to no h6: type of investigation code updated to 3331: analysis of production records, h6: type of investigation code updated to 4111: communication/interviews, h6: type of investigation code updated to 4110: trend analysis, h6: type of investigation code updated to 4114: device not returned, h6: type of investigation code updated to 4117: device not accessible for testing, h6: type of investigation code updated to 4109: historical data analysis, h6: investigation findings code updated to 213: no device problem found, h6: investigation conclusions code updated to 4315: cause not established, h10: additional narratives/data.
|
| |
|
Search Alerts/Recalls
|
|
|