| Model Number |
13117545 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2021 a patient underwent a extreme lateral interbody fusion procedure from l3 to l5 fixed with posterior fixation.On a unknown date it was discovered that the right screw at l5 was deviated inward.This deviation was confirmed to be the result of proper trajectory not being achieved in the index procedure.On (b)(6) 2021 a revision occurred where the deviated screw was re used and put in the proper position.The was no report of any adverse consequences to the patient as a result of the malplaced screw and the patient is considered to be recovering.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot number was reported to be either n288197 or n295595.No product was returned as no product fault was alleged and the involved device remains in-situ.No radiographs could be provide showing the initial or revised construct confirming the complaint.No additional information could be obtained.The root cause is considered to be an inadvertent use error and the result of screw malplacement during the index procedure.Labeling review: "potential adverse events and complications as with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries." "warnings, cautions and precautions: the implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure.".
|
| |
|
Search Alerts/Recalls
|