• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REDAPT SLVD MONO STEM 300MM SZ 19 SO; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REDAPT SLVD MONO STEM 300MM SZ 19 SO; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Model Number 71354844
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hematoma (1884); Unspecified Infection (1930)
Date of Event 09/28/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, after a medically indicated revision of right hip tha performed on (b)(6) 2020, where redapt slvd mono stem 300mm sz 19 so was implanted, a second revision surgery was necessary due to an unspecified infection.An hematoma was also reported to be drained.Additional details about primary surgery and the patient's final outcome are unknown.This de-identified clinical data is related to a post market clinical follow-up activity for redapt sleeved mono stems.As the data collection activity has been done retrospectively and data is provided anonymized, the information provided is limited and further information cannot be obtained.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, no relevant supporting clinical information could be provided to assist with this clinical investigation as the data collected from the post market clinical follow-up was anonymous.Without clinically relevant patient-specific supporting documentation, a thorough medical investigation cannot be performed.The patient's current condition is unknown and the patient impact beyond the reported events could not be determined based on the limited information provided.Should any additional clinical information be provided this complaint will be re-evaluated.A review of complaint history did not reveal additional complaints for the listed device for the same failure mode.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but are not limited to contamination, patient reaction, and post-operative healing issue.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.D10: add concomitants.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDAPT SLVD MONO STEM 300MM SZ 19 SO
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12524189
Report Number1020279-2021-07119
Device Sequence Number972694
Product Code LZO
UDI-Device Identifier00885556580295
UDI-Public00885556580295
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K162303
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,study
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71354844
Device Catalogue Number71354844
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/01/2021
Supplement Date Received by Manufacturer10/22/2021
Initial Report FDA Received Date09/24/2021
Supplement Report FDA Received Date10/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
REDAPT MOD SLV MEDSZ 18-19 STIKTITE W/HA/UNK
Outcome Attributed to Adverse Event Hospitalization; Other;
Patient Age57 YR
-
-