| Model Number |
9735542 |
| Medical Device Problem Code |
Unexpected Therapeutic Results (1631)
|
| Health Effect - Clinical Code |
Unspecified Tissue Injury (4559)
|
| Date of Event |
09/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient age not available from the site.Patient weight not available from the site.No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a thermal therapy system being used in an electrode and probe placement procedure.It was reported that the fibers were not changed during a two fiber case.The energy was put higher than intended during the test ablation and a small unwanted lesion was created.The patient was okay.There were no severe consequences.The fiber was used for the case and worked fine.A new fiber was inserted and a scan was taken to look at the created ablation.
|
| |
|
Event or Problem Description
|
|
Additional information was received stating that there was a 10 minute delay in the case.A small ablation at the tip of the first pulled back catheter.Close to the internal capsula.No consequences for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
H2) additional information was added to the event description.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|