• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX; TUBE, TRACHEOSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX; TUBE, TRACHEOSTOMY Back to Search Results
Model Number 101/860/080CZ
Device Problem Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2021
Event Type  malfunction  
Manufacturer Narrative
One smiths medical tracheostomy|pvc - portex tubes bluselect was returned for analysis in a used condition.Under visual inspection the sample appeared to be in good condition.During manufacturing process, the devices are 100% inflation tested, which includes inflating each device cuff and leaving for a 12-hour period.Reductions in pressure over this time are considered a failure and the device would be rejected.The inflation test was repeated on the received samples.It was confirmed that after 12 hours the cuff was still fully inflated.Also, no leak was found when inflated device was put under water.Based on the evidence, the complaint was not confirmed, and no fault was found.
 
Event Description
Information was received indicating that while in use of a smiths medical tracheostomy|pvc - portex tubes bluselect, air pressure of the cuff was often unable to be maintained.No patient consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX
Type of Device
TUBE, TRACHEOSTOMY
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
MDR Report Key12525296
MDR Text Key273085953
Report Number3012307300-2021-09579
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076158
UDI-Public15019517076158
Combination Product (y/n)N
PMA/PMN Number
K173384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 09/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101/860/080CZ
Device Catalogue Number101/860/080CZ
Device Lot Number3903847
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2021
Was the Report Sent to FDA? No
Date Manufacturer Received09/07/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-