| Model Number |
AIRLIFE¿ ADULT MANUAL RESUSCITATOR, 40'' (1.0M) OXYGEN RESERVOIR TUBING, ADULT M |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Pulmonary Infarction (2021)
|
| Date of Event |
08/23/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At this time, the suspect device has not been returned for evaluation.At this time, no root cause could be determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported that the 2k8004 resus, adlt w/mask, resv bag reservoir on the ambu bags was wrapped at the distal end opening where the tubing goes through the reservoir, it was folded over blocking air from escaping causing excessive pressures which resulted in a pneumothorax with one patient.
|
| |
|