| Catalog Number |
320617 |
| Medical Device Problem Code |
Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.Investigation summary: exec summary: no samples were returned therefore bd was not able to duplicate or confirm the customer¿s indicated failure and the root cause is undetermined.A review of the complaint lot history check was performed and this is the 1st related complaint for needle clog on this lot #.No non-conformance's were raised in association with this type of event for this lot concluding all inspections were performed as per the applicable operations and met qc specifications.Capa/sa: based on the above no additional investigation and no corrective or preventative action (capa) or situational analysis (sa) are required at this time.Dhr review: a lot history review was carried out and no related non conformances were raised in association with this packaged lot concluding all inspections were performed as per the applicable operations and met qc specifications.
|
| |
|
Event or Problem Description
|
|
It was reported that 1 bd pn relion 31g x 6mm was unable to deliver insulin or medication.The following information was provided by the initial reporter: the consumer reported that the patient end of needle was missing and there was no insulin flow during injection.Date of event: unknown.Samples: discarded.
|
| |
|