Model Number 0046140
Medical Device Problem Codes Biocompatibility (2886); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Pain (1994); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that there was a malfunction of nasogastric tube.Per follow up received on 30aug2021 the customer stated that on one patient the nasogastric tube leaked the gastric content where the anti reflux port meets the gastric port and another patient there was a continuous whistling sound that came from the same area.It was also stated that the tube removed from the patient and experienced trauma of having another one inserted with the same issue where the placement was confirmed for each insertion.The tube was ultimately removed the second time and the patient refused another insertion which delayed the healing related to the diagnosis.No medical intervention was reported.
 
Additional Manufacturer Narrative
Upon further review, bd has determined that this event is not reportable.
 
Event or Problem Description
It was reported that there was a malfunction of nasogastric tube.Per follow up received on 30aug2021 the customer stated that on one patient the nasogastric tube leaked the gastric content where the anti reflux port meets the gastric port and another patient there was a continuous whistling sound that came from the same area.It was also stated that the tube removed from the patient and experienced trauma of having another one inserted with the same issue where the placement was confirmed for each insertion.The tube was ultimately removed the second time and the patient refused another insertion which delayed the healing related to the diagnosis.No medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® NASOGASTRIC SUMP TUBE WITH PREVENT® ANTI-REFLUX FILTER
Common Device Name
NASOGASTRIC TUBE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12525678
Report Number1018233-2021-05935
Device Sequence Number13696283
Product Code FEG
UDI-Device Identifier00801741052316
UDI-Public(01)00801741052316
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K960176
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model Number0046140
Device Catalogue Number0046140
Device Lot NumberNGFP1769
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/30/2021
Supplement Date Received by Manufacturer05/24/2022
Initial Report FDA Received Date09/24/2021
Supplement Report FDA Received Date05/27/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-