Device was returned for additional evaluation and investigation.A review of the device history record, which includes verification of all steps in the manufacturing of the catheter kit, verification of sterilization, and packaging for subject catheter kit was performed.The review did not identify any non-conformances, issues or capas associated with catheter kit.Additional physical investigation was performed on the device, but the alleged issue could not be confirmed.Two sections of catheter were returned.The first section was 39.37 cm long with the distal end.There was a barb connection at 36.2 cm from the distal tip.The second section was 36.2 cm long, with no distal end.The sections were found to be patent with air and swi.No leaks were observed in either section of catheter, and the complaint could not be confirmed.It is possible that the catheter was cut at the observed leak, but that cannot be determined from the returned sections of catheter or the information in the complaint.Per the instructions for use of the device, catheter leakage is a known possible risk of use of the device.Internal complaint number: (b)(4).
|