• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC AORTIC CUTTERS, 5-PACK (3.8MM); CUTTER, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC AORTIC CUTTERS, 5-PACK (3.8MM); CUTTER, SURGICAL Back to Search Results
Model Number AORTIC CUTTERS, 5-PACK (3.8MM)
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Trackwise id # (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
 
Event or Problem Description
The hospital reported that during a coronary artery bypass procedure using aortic cutters (3.8mm).They pressed the cutter operation button after releasing the safety lock, but the button was hard and the cutter did not come out.There was no problem when they prepared and used a substitute cutter.
 
Additional Manufacturer Narrative
Trackwise # (b)(4).Analysis of production: (3331/213/67) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213/67) the review of the historical data indicates that no other similar complaints was reported for the same lot number and reported failure mode.Trend analysis: (4110/213/67) the overall 24 month product complaint trend data for the period oct 2019 through sept 2021 was reviewed.There were no triggers identified for the review period.Communication/interviews: (4111/213/67) communication/interviews were performed to obtain all possible information.
 
Event or Problem Description
The hospital reported that during a coronary artery bypass procedure using aortic cutters (3.8mm).They pressed the cutter operation button after releasing the safety lock, but the button was hard and the cutter did not come out.There was no problem when they prepared and used a substitute cutter.It has no effect on the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC CUTTERS, 5-PACK (3.8MM)
Common Device Name
CUTTER, SURGICAL
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key12526141
Report Number2242352-2021-00761
Device Sequence Number1547841
Product Code FZT
UDI-Device Identifier00607567700260
UDI-Public00607567700260
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/13/2022
Device Model NumberAORTIC CUTTERS, 5-PACK (3.8MM)
Device Catalogue NumberAC-3038
Device Lot Number25157565
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2021
Supplement Date Received by Manufacturer12/07/2021
Initial Report FDA Received Date09/24/2021
Supplement Report FDA Received Date12/07/2021
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-