Siemens has evaluated the card lot performance: the in-house performance for the card lot in question, 01-21177-50, was investigated for po2 performance and did not identify any product deficiencies.Card lot 01-21177-50 was tested with arterialized blood with a target po2 of approximately 80 mmhg and with aqueous control fluids eurotrol l1 and eurotrol l3 at the time of product release.Aqueous fluids and arterialized blood displayed no indication of discrepant results at the time of release.There were no other complaints received on card 01-21177-50.Therefore, there is no evidence that the system or reagent cards are not performing as intended.Siemens recommends using the correct reference ranges associated with the sample type collected.According to the data, all of the po2 results were acceptable results for the earlobe capillary sample type used for analysis.Additionally, the customer was using earlobe capillary samples.According to clsi c46-a blood gas and ph analysis and related measurements: approved guideline, arterial blood samples are preferred for blood gas analysis.
|