• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) CR2 DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) CR2 DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number CR2
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2021
Event Type  malfunction  
Manufacturer Narrative
Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event Description
The customer contacted physio-control to report a non-critical issue with their device.There was no patient use associated with the reported event.Upon further investigation it was found that device was missing the magnetic pin the lid.In this state the device will not turn on automatically which can lead to a use error resulting in a failure to deliver defibrillation.
 
Manufacturer Narrative
The customer has been provided with a replacement device.The cause of the missing lid magnet was isolated to the lid assembly.An engineering investigation has determined that due to the design of the lcpr2 lid switch, absence of the lid magnet allows current to flow from the battery, even when the device is in standby mode.This will reduce the life of the battery.Therefore, the reported issue, the loss of lid/lid magnet, is associated with two hazardous situations: opening the lid does not turn on the device, and, higher than intended current draw while the device is in standby mode results in a prematurely depleted the battery.Replacement of the device lid so that the magnet is present in the device allows the lid switch to function as designed; turning on/off the device when the lid is opened and preventing current draw when the device lid is closed.The lpcr2 operating instructions has been updated to include additional troubleshooting tips to direct the customer to act when device behavior indicates the lid magnet may be missing.A new warning informs the customer of the risk associated with a missing magnet.
 
Event Description
The customer contacted physio-control to report a non-critical issue with their device.There was no patient use associated with the reported event.Upon further investigation it was found that device was missing the magnetic pin the lid.In this state the device will not turn on automatically which can lead to a use error resulting in a failure to deliver defibrillation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) CR2 DEFIBRILLATOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key12526260
MDR Text Key273124590
Report Number0003015876-2021-01837
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
P170018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Biomedical Engineer
Remedial Action Recall
Type of Report Initial,Followup
Report Date 11/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberCR2
Device Catalogue Number99512-001011
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/05/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/30/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number3015876-01/14/2021-001-C
Patient Sequence Number1
-
-