MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 97715
Device Problems Overheating of Device (1437); Communication or Transmission Problem (2896)
Patient Problem Burn(s) (1757)
Event Date 09/01/2021
Event Type  malfunction  
Manufacturer Narrative
Continuation of concomitant products: product id 97755.Lot# serial# (b)(4).Implanted: explanted: product type recharger.Information references the main component of the system.Other relevant device(s) are: product id: 97755, serial/lot #: (b)(4), ubd: , udi#: (b)(4).Date is approximate.
 
Event Description
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for non-malignant pain.It was reported that they were having issues with consistently communicating with the stimulator when using the recharger.The patient stated they charge once a day or every other day, a few mins "here and there", and the stimulator is currently at 30%.The patient stated they tried moving the paddle of the recharger around and using passive recharge mode but the issue would not resolve.The patient stated the recharger, where the paddle and cord meet and at the relay box, would get "fire hot" and burned the patient's skin.The patient stated they put electrical tape around the cord, but the patient was still having issues with the recharger.An email was sent to the repair department to replace the recharger.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key12526268
MDR Text Key273123272
Report Number3004209178-2021-14371
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781702
UDI-Public00643169781702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Date Manufacturer Received09/23/2021
Date Device Manufactured03/09/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age32 YR
-
-