| Model Number |
E360T-SY-IN |
| Medical Device Problem Code |
Smoking (1585)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: medtronic conducted an investigation based upon all information received.It was reported that smoke was observed coming from the e360 ventilator.The device was made available for evaluation.The service personnel (sp) evaluated and could not confirm the reported issue as the ventilator and air compressor were turned on normally.Sp additionally found a tidal volume mismatched.Performed calibration successfully.All the parameters were checked and verified and functioned properly.The reported issue could not be confirmed.The most likely cause could not be established from the information available.There was no evidence of smoke damage or any smell.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all medtronic quality specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that smoke was observed coming from the e360 ventilator.The ventilator was not in use on a patient at the time of the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to section h6 evaluation code result and conclusion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|