• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL ARTHSCO,4/140_30_QIK/STRKR HUB; RIGID ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL ARTHSCO,4/140_30_QIK/STRKR HUB; RIGID ENDOSCOPE Back to Search Results
Model Number 242043
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Investigation summary: the complaint device was received at the service center and evaluated.It was reported that bent scope.Per service reports, this complaint can be confirmed.During the service evaluation the following defects were identified: minor scratches on the unit.Distal tip damaged, shaved down and bent.Objective damaged and dented.Window is shaved down.Optics are damaged.Image is cloudy/foggy.Tube bent and dented.The objective and objective window were replaced, internal scope cleaning, laser weld eyepiece as originally designed and refibering to resolve the issues.After repair, the device was found to be working according to the specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event or Problem Description
It was reported by the sales rep that during a shoulder repair procedure on (b)(6) 2021, it was observed that the scope on the arthsco,4/140_30_qik/strkr hub device was bent.During in-house engineering evaluation, it was determined that the image on the device was cloudy/foggy.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTHSCO,4/140_30_QIK/STRKR HUB
Common Device Name
RIGID ENDOSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key12526737
Report Number1221934-2021-02790
Device Sequence Number4427938
Product Code HRX
UDI-Device Identifier10886705026234
UDI-Public10886705026234
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K971996
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number242043
Device Catalogue Number242043
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2021
Initial Report FDA Received Date09/24/2021
Patient Sequence Number1
-
-