Merge cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information from merge healthcare and other vendors systems including images, hemodynamic studies and reports, measurements (via import from dicom structured reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.Merge cardio is intended to allow users to review diagnostic and non-diagnostic quality images, annotate studies, perform digital subtraction on images, to perform quantitative measurements on images (including but not limited to quantitative coronary analysis, left ventricular analysis, time, area, length, velocity, angle, volume, and velocity-time integrals), to generate physician generated clinical reports (via structure reporting and template based tools), and to store this information in a database.On (b)(6) 2021, a customer contacted merge healthcare to report that two reports were not available for viewing.Merge healthcare support found that the reports were never confirmed but were in confirmed status.The logs had been truncated so they were unavailable for troubleshooting purposes.Support manually reset the status of the reports and they were then able to be read and confirmed by the physician.No further action is necessary at this time.This has the potential to delay patient treatment and/or diagnosis.There have been no reports of patient injury or harm as a result of this issue.Reference complaint (b)(4).
|
|
This supplemental report is submitted to the fda in accord with applicable regulations and as indicated by merge healthcare in the initial report submitted 9/24/2021.As of 9/28/2021, merge healthcare is ceasing reporting on merge cardio report status sync issues.The issue in question presents when a report is in a confirmed status in the database but was not confirmed by the physician.This issue is caused by network issues and not caused by or contributed to by merge cardio in any way.Review of this issue based on current knowledge of the device, clinical workflow, existing mitigations against user errors, and the presence of no death or injury associated with the issue at any point between 2016 and 2021 indicates that the device is operating as designed and that this issue presents remote likelihood of injury or harm to a patient.No further action is required.Revised information contained in this supplemental report includes the following: g3 - date new information received by manufacturer.G6 - indication that this is follow-up report 001.H1 - indication of malfunction as reportable event.H2 - indication of additional information.H10 - additional narrative/data.
|