| Model Number |
0036280 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device in question has been received by conmed and has entered into the evaluation process.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
During incoming inspection, the distributor rejected this device, catalog # 0036280, lot 202006265, for a possible breach of sterility due to an insufficient heatseal.There was no contact with any patient as this was found during incoming inspection in (b)(6).The completion of the device evaluation confirmed an insufficient heatseal; therefore, this complaint meets the criteria for a reportable event.Due to the breach in sterility, this report is being raised as a malfunction with potential for injury upon reoccurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: investigation of the customer's complaint of an insufficient heat seal leading to a breach in sterility has confirmed the reported issue.Conmed received two 0036280 returned unopened in original packaging.The lot number of the devices, 202006265, was verified.Open holes were found on the packaging of both devices.A functional inspection was not needed as the holes were obvious failures / breach in sterility.The manufacturing documents from the device history record have been reviewed and found no abnormalities that would contribute to this issue.A two-year lot history review found this is the only complaint for this lot number and failure mode.A two-year review of complaint history revealed there has been a total of 27 complaints, regarding 39 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.(b)(4).Conmed encourages the inspection and/or test of all medical equipment prior to use.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
During incoming inspection, the distributor rejected this device, catalog # 0036280, lot 202006265, for a possible breach of sterility due to an insufficient heatseal.There was no contact with any patient as this was found during incoming inspection in japan.The completion of the device evaluation confirmed an insufficient heatseal; therefore, this complaint meets the criteria for a reportable event.Due to the breach in sterility, this report is being raised as a malfunction with potential for injury upon reoccurrence.
|
| |
|