This complaint was reopened in reference to (b)(4).Manufacturing site evaluation: manufacturing and quality control data.The progav® was manufactured by a qualified employee in may 2013.Deviations during assembly did not occur.The valve was sterilized by miethke and released for shipment after final inspection.The progav® has a normal pressure range of 0 to 20 cmh2o.The valve was inspected as article fv413t.All parameters have been inspected and approved during the manufacturing process.All parameters have been assessed as meeting specification.Device examination: the valve was received submersed in an unidentified liquid in the provided returnkit.Visual inspection: no significant deformations or damage to the valve were detected during the visual inspection.Permeability test: a permeability test has shown that the valve was permeable.Adjustment test: the progav® was tested and was not adjustable throughout the normal range.Braking force and brake function test: the brake functionality test has shown that the brake function was operational, however, the braking force was not able to be measured due to the non-adjustability of the valve.Result: first, a visual inspection of the progav® was performed.No significant deformations or damage to the valve were detected during the visual inspection.Next, the permeability, adjustability, as well as the brake functionally and brake force, of the valve were tested.The valve was permeable, but was not able to be adjusted.Finally, the valve was dismantled.A build-up of a substance (likely protein) was observed inside the valve.Based on the investigation, the claim of valve occlusion was confirmed.This was likely due to the deposits observed inside the valve.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hydrocephalus (hc)-therapy by shunt implants.Manufacturing defects at the time of release have been excluded.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required.
|