Model Number 3425-1380S
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 08/27/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device retained by hospital.
 
Event or Problem Description
The customer reported there was a spontaneous fracture of the implantable medical device.Revision surgery was required to put the nail back in place.
 
Additional Manufacturer Narrative
Please note correction to d4 lot number.The reported event could be confirmed based on images provided, only.A physical device inspection was not possible since the affected device was not made available for evaluation and thus, inspection was limited to the images available.Based on these images a broken nail was verified.The nail is completely broken in the webs of the proximal drill hole for the lag screw.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history did not indicate any abnormalities.No corrective actions are required at this time.The file will be closed formally in accordance to our procedures.In case the item and / or substantive information will become available in future the file will be reviewed and reopened.
 
Event or Problem Description
The customer reported there was a spontaneous fracture of the implantable medical device.Revision surgery was required to put the nail back in place.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø11X380MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
marilyne chaumont
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key12530129
Report Number0009610622-2021-00723
Device Sequence Number5169248
Product Code HSB
UDI-Device Identifier07613252273844
UDI-Public07613252273844
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K034002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 03/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/31/2022
Device Model Number3425-1380S
Device Catalogue Number34251380S
Device Lot NumberK046377
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/03/2021
Supplement Date Received by Manufacturer02/08/2022
Initial Report FDA Received Date09/24/2021
Supplement Report FDA Received Date03/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/15/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-