• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. CORFLO NASOGASTRIC/NASOINTESTINAL NEONATAL FEEDING TUBE WITH ENFIT CONNECTOR; DH CPK NG TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. CORFLO NASOGASTRIC/NASOINTESTINAL NEONATAL FEEDING TUBE WITH ENFIT CONNECTOR; DH CPK NG TUBES Back to Search Results
Model Number 41-1155
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Discomfort (2330)
Date of Event 08/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device history record is not possible as no lot number was provided.The actual complaint product was returned for evaluation.The root cause of the reported issue could not be conclusively determined.However; this seems to be a user related problem since as per the ifu, vigorous syringe force should not be used to irrigate, administer liquids, or unblock the tube.All information reasonably known as of 24 sep 2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc.Complaint database and identified as complaint (b)(4).
 
Event or Problem Description
It was reported that the ng [nasogastric] tube was "broken." according to the report, the "patient [was] fussy all day with feeds, kub [kidney-ureter-bladder x-ray] obtained which showed that ngt was broken in patient's stomach (distal tip hanging on by a thread).Ngt pulled and replaced." no injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORFLO NASOGASTRIC/NASOINTESTINAL NEONATAL FEEDING TUBE WITH ENFIT CONNECTOR
Common Device Name
DH CPK NG TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
MDR Report Key12530352
Report Number9611594-2021-00137
Device Sequence Number3990752
Product Code KNT
UDI-Device Identifier00350770460604
UDI-Public00350770460604
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K831328
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number41-1155
Device Catalogue NumberN/A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2021
Initial Report FDA Received Date09/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-