• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION POWERFLEXPRO 8MM4CM 80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION POWERFLEXPRO 8MM4CM 80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 4400804S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, an 8mm x 4cm 80 powerflex pro percutaneous transluminal angioplasty (pta) balloon catheter was used; however, it ruptured.There was difficulty crossing the lesion.There was no reported patient injury.This was for an endovascular therapy (evt) case.There was no difficulty removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks nor other damages noted prior to inserting the product into the patient.The device was prepped normally (i.E., maintain negative pressure).The same indeflator was used successfully with other devices.There was no resistance/friction while inserting the balloon through the rotating hemostatic valve, or the guide catheter.There was no difficulty advancing the balloon catheter through the vessel.The catheter was in an acute bend.The plunger depressed into the syringe/indeflator when trying to inflate.Leakage was noted from either the balloon, shaft, hub, or other unknown segment.The balloon catheter was not kinked while being used.The balloon catheter was removed easily.The product was removed intact (in one piece) from the patient.A new device was used, and the procedure was completed.The device will not be returned for evaluation as it was discarded by mistake.
 
Additional Manufacturer Narrative
Complaint conclusion: as reported, an 8mm x 4cm 80 powerflex pro percutaneous transluminal angioplasty (pta) balloon catheter was used; however, it ruptured.There was difficulty crossing the lesion.There was no reported patient injury.This was for an endovascular therapy (evt) case.There was no difficulty removing the product from the hoop, the protective balloon cover, the stylet, or any of the sterile packaging components.There were no kinks nor other damages noted prior to inserting the product into the patient.The device was prepped normally (i.E., maintain negative pressure).The same indeflator was used successfully with other devices.There was no resistance/friction while inserting the balloon through the rotating hemostatic valve, or the guide catheter.There was no difficulty advancing the balloon catheter through the vessel.The catheter was in an acute bend.The plunger depressed into the syringe/indeflator when trying to inflate.Leakage was noted from either the balloon, shaft, hub, or other unknown segment.The balloon catheter was not kinked while being used.The balloon catheter was removed easily.The product was removed intact (in one piece) from the patient.A new device was used, and the procedure was completed.The device was not returned for evaluation as it was discarded by mistake.A product history record (phr) review of lot 82220677 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics, although unknown, may have contributed to the reported event.However, with the limited amount of information provided regarding vessel characteristics and without the return of the device for analysis, it is difficult to draw a clinical conclusion between the device and the event reported.According to the safety information in the instructions for use, ¿if resistance is met during manipulation, determine the cause of the resistance before proceeding.Balloon pressure should not exceed the rated burst pressure.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence), will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over pressurization.Use only the recommended balloon inflation medium.Never use air or any gaseous medium to inflate the balloon.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERFLEXPRO 8MM4CM 80
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
MDR Report Key12531641
Report Number9616099-2021-04897
Device Sequence Number1428816
Product Code LIT
UDI-Device Identifier20705032059814
UDI-Public20705032059814
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112797
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model Number4400804S
Device Catalogue Number4400804S
Device Lot Number82220677
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/02/2021
Supplement Date Received by Manufacturer09/28/2021
Initial Report FDA Received Date09/27/2021
Supplement Report FDA Received Date10/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-