Model Number N/A
Medical Device Problem Codes Break (1069); Fracture (1260); Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Initial report.Report source, foreign - event occurred in (b)(6).Concomitant medical products: it has been indicated that the product is in process of being returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that during impaction of the femur, the plastic of the impactor broke off and has to be replaced.Nothing fell into the patient and no harm was reported.Surgery completed without delay.
 
Event or Problem Description
It was reported that during impaction of the femur, the plastic of the impactor broke off and has had to be replaced.Nothing fell into the patient and no harm was reported.Surgery completed without delay.
 
Additional Manufacturer Narrative
(b)(4).This final report is being submitted to relay additional information.No product was returned; visual and dimensional evaluations could not be performed.Lot identification is necessary for review of device history records, which was not provided.A review of complaint history was assessed for three years prior to the notification date and identified (8) similar complaints for item #32-420127 (including initiating complaint).A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any additional information becomes available, then the complaint will be reopened and further investigated.H3 other text: product not returned.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: b4, b5, d4, g3, g6, h1, h2, h3, h6, h10.Visual examination of the returned product identified deformation at the edges of the impaction area, numerous scratches and gouges over the surface of the impaction area and small fractures at the edge of impaction surface.The instrument has potentially been in the field for approximately 3 years and 1 month.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Device is used for treatment.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF UNI PH3 FEMORAL IMPACTOR
Common Device Name
KNEE ARTHROPLASTY
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12532044
Report Number3002806535-2021-00403
Device Sequence Number1350723
Product Code NRA
UDI-Device Identifier05019279477333
UDI-Public(01)05019279477333(11)190902(10)Z19G1770
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number32-420127
Device Lot NumberZ19G1770
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2021
Supplement Date Received by Manufacturer09/21/2021
10/10/2022
Initial Report FDA Received Date09/27/2021
Supplement Report FDA Received Date12/23/2021
10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-