Model Number 2P56-22
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.Reporter phone complete entry = (b)(6).Reporter postal code had leading zero added to meet minimum field requirement.
 
Event or Problem Description
The customer observed a falsely elevated lactate dehydrogenase (ldh) result for one patient on an architect c8000 analyzer.The following data was provided (customer¿s reference range is 125-220 u/l): initial result was 491, repeats were 200 and 212 u/l.There was no impact to patient management reported.
 
Additional Manufacturer Narrative
The complaint investigation for a falsely elevated lactate dehydrogenase (ldh) result included a search for similar complaints, the review of complaint text, trending data, labeling, and device history records.Trending review determined no related trends for the product.Return testing was not completed as returns were not available.Device history record review did not identify any non-conformances or deviations with the likely cause list number and complaint issue.Manufacturing documentation for the likely cause list number was reviewed and did not identify any issues.Labeling was reviewed and sufficiently addresses the customer's issue.Based on the information provided and abbott diagnostics¿ complaint investigation, no systemic issue or deficiency of lactate dehydrogenase (ldh) assay, list number 02p56-22, was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LACTATE DEHYDROGENASE (LDH)
Common Device Name
NAD REDUCTION/NADH OXIDATION, LACTATE DEHYDROGENASE
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
MDR Report Key12532492
Report Number3002809144-2021-00581
Device Sequence Number19117827
Product Code CFJ
UDI-Device Identifier00380740161460
UDI-Public00380740161460
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2P56-22
Device Catalogue Number02P56-22
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/14/2021
Supplement Date Received by Manufacturer10/11/2021
Initial Report FDA Received Date09/27/2021
Supplement Report FDA Received Date10/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC C8 PROC MOD, 01G06-11, C803904; ARC C8 PROC MOD, 01G06-11, C803904
-
-