SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 56MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, Back to Search Results
Model Number 71335556
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Metal Related Pathology (4530); Nodule (4551)
Date of Event 07/02/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal reference number: (b)(4).
 
Event or Problem Description
''(b)(6) legal'': it was reported that after a r3 tha construct was implanted on (b)(6) 2008 the plaintiff experienced elevated metal ion levels on blood and pseudotumor.A revision surgery was performed on (b)(6) 2019 to treat these adverse events.During the surgery, a huge amount of brownish-black fluid was found on the hip.Additionally, grayish-black tissue in a large portion of the hip was found, and a pseudotumor was removed around the acetabulum and femoral portion on the hip.Microscopy study of the tissue removed showed tissue necrosis, fibrosis, chronic inflammation, with a good foreign body histiocyte reaction.Patient outcome is unknown.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, although alval was suspected, the pathology report did not confirm.The elevated metal ion levels and intraoperative findings of blackish/brownish liquid, black gray caseous fabrics, and pseudotumor along with the synovial tissue fragments that show necrosis, fibrosis, chronic inflammation and histiocytic reaction to foreign body may be consistent with an adverse reaction to metal debris.However, based on the information provided, the clinical root cause of the clinical reactions cannot be confirmed.It cannot be concluded that the clinical reactions were associated with a malperformance of the implant.The patient impact beyond the revision cannot be determined.No further clinical assessment is warranted at this time.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but are not limited to traumatic injury, joint tightness, friction or wear.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE ACET SHELL 56MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL,
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12532643
Report Number1020279-2021-07176
Device Sequence Number16497408
Product Code MRA
UDI-Device Identifier03596010598264
UDI-Public03596010598264
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K070756
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/03/2017
Device Model Number71335556
Device Catalogue Number71335556
Device Lot Number07LM04180
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/03/2021
Supplement Date Received by Manufacturer10/27/2021
Initial Report FDA Received Date09/27/2021
Supplement Report FDA Received Date10/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MODULAR SLEEVE +4MM 12/14/07KW14071; SYN POR HO FEM COM SZ 18/ 02JM02565; MODULAR SLEEVE +4MM 12/14/07KW14071; SYN POR HO FEM COM SZ 18/ 02JM02565
Outcome Attributed to Adverse Event Required Intervention;
-
-