| Model Number |
8888145002 |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, post insertion, during dialysis session, no blood flow observed from the blue (venous) port.The product was replaced with another lot.The catheter was not repaired.There was no leak.Betadine was the cleaning agent used on the device.Tego was not utilized and there was no luer adapter issue.There was no patient injury.
|
| |
|
Event or Problem Description
|
|
According to the reporter, 5 days after catheter was inserted, during dialysis session, no blood flow was observed from the blue (venous) port.The customer tried to reverse the lines, but reverse flow was not successful and it was not much of a good flow.The product was replaced with the same product id (identifier) and same lot on the next day of the event.The new device was used successfully and the procedure was completed.Nothing unusual observed on the device prior to use, flushing was done prior to use and it was okay.The line was not hard to flush with the syringe, there was blood return prior to syringe flush, heparin was the anti-coagulation used, and this was not the infusion line.No other products being utilized with the device.There was no problem with the catheter's dimension, the dimension(s) of the catheter matched what was indicated on the label, there was no occlusion, no other defects/d amages found on the product, and the catheter was not repaired.There was no leak.Betadine was the cleaning agent used on the device.Tego was not utilized and there was no luer adapter issue.There was no blood loss.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, d6a, d6b, d9, g1 (mfr contact first name, last name, email, phone number), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that there was an insufficient flow issue.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|