H2: h3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A clinical review states this case reports the broken jaw of the fristpass suture passer was removed from inside of the patient with arthroscopic graspers.Based on the limited information provided the root cause of the reported jaw breakage could not be determined.According to report the procedure was completed with a non-significant delay using a back-up device.Since there was no patient injury or other complications reported, no further clinical/medical assessment is warranted at this time.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|