• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CUSTOM-PAK SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CUSTOM-PAK SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number ASKU
Device Problem Material Fragmentation (1261)
Patient Problems Foreign Body In Patient (2687); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  Injury  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported a microfibrile detached from a gown and was retained in a patient's eye during a cataract procedure, was removed on the following day.
 
Manufacturer Narrative
The customer did not retain the product lot information for this custom pak, therefore the device history records traceable to the reported procedure pack could not be reviewed.The customer reported that fibers were coming detached from the gown in their pak.No sample has been returned for evaluation; therefore, the condition of the product could not be verified.When this type of issue occurs, a sample will need to be returned so that a proper investigation can be conducted.If possible, the customer should provide company with a photo of the reported foreign material in order to help with the investigation of the issue.The root cause of the customer's complaint could not be established as a sample has not been received and the condition of the product could not be verified.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.Custom paks are manufactured in an environmentally controlled area (eca) that is routinely monitored.In addition, the manufacturing area operates in a controlled positive pressure environment and is continuously monitored to prevent entry of foreign material and hair.No action will be pursued at this time for this occurrence.Complaints are continuously monitored to identify product and/or process areas where debris and hair is occurring.In addition, routine training and awareness is conducted with all manufacturing associates to reinforce the importance of wearing proper ppe (personal protective equipment) and maintaining clean work areas.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM-PAK SURGICAL PROCEDURE PACK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key12536290
MDR Text Key273538600
Report Number1644019-2021-00591
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/31/2021
Initial Date FDA Received09/27/2021
Supplement Dates Manufacturer Received11/24/2021
Supplement Dates FDA Received12/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-