• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROCLAIM 5 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROCLAIM 5 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG Back to Search Results
Model Number 3660
Device Problem Wireless Communication Problem (3283)
Patient Problem Insufficient Information (4580)
Event Date 09/09/2021
Event Type  malfunction  
Event Description
It was reported the patient was unable to communicate with the ipg following an unrelated surgical procedure.Next course of action is undetermined at this time.
 
Manufacturer Narrative
An inoperable implant was reported to abbott.It is unknown if the system was set to surgery mode while the patient underwent an unrelated surgery where electro-cautery may have been used.The implant was not returned for analysis.A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.Based on the information received, the cause of the reported incident is consistent with user error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 5 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
MDR Report Key12537117
MDR Text Key273537683
Report Number3006705815-2021-04772
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067031419
UDI-Public05415067031419
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2022
Device Model Number3660
Device Catalogue Number3660
Device Lot NumberA000096915
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-