• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: O3 REGIONAL OXIMETER MODULE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

O3 REGIONAL OXIMETER MODULE Back to Search Results
Model Number 24667
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product has been returned to the local facility but has not yet been received at the main office for evaluation.Once returned and investigated, a follow-up report will be submitted.
 
Event or Problem Description
The customer reported the device shows impossible values.No patient impact or consequences were reported.
 
Additional Manufacturer Narrative
Additional manufacturing narrative: the returned device was evaluated.During evaluation the device passed all visual and functional testing.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed.
 
Event or Problem Description
The customer reported the device shows impossible values.No patient impact or consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O3 REGIONAL OXIMETER MODULE
Common Device Name
OXIMETER
MDR Report Key12537481
Report Number3011353843-2021-00255
Device Sequence Number3431809
Product Code MUD
UDI-Device Identifier00843997010771
UDI-Public00843997010771
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K033296
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 09/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24667
Device Catalogue Number9637
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/06/2021
Supplement Date Received by Manufacturer10/05/2021
Initial Report FDA Received Date09/27/2021
Supplement Report FDA Received Date10/29/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/11/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
-
-