| Model Number |
324912 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
08/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no capa/sa is required at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that syringe 1.0ml 31ga 6mm 10bag 500cs scale markings were misaligned.The following information was provided by the initial reporter: it was reported that insulin syringes has the needle shield off the syringe in bag.Caused a clean needle stick to employee that was treated with minor first aid.Verbatim: minor first aid was given to our employee.This was the first incident reporting of this employee.We have had four bags with insulin syringes that had at least one cap off in the plastic bag.(these are in our sharps container.) we have come across some insulin syringes that the caps are off in the plastic bag exposing the needle.This is a safety issue as we had an employee puncture his thumb which is then a biohazard.
|
| |
|