ZIMMER MANUFACTURING B.V. SHELL 50 MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/30/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Foreign: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0002648920-2021-00290.
 
Event Description
It has been reported that, during a surgery, the shell and the liner of the size 50 did not connect.The size 49 were finally implanted.No additional information.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product identified one of two tabs of the lock ring was seized in the lock ring groove of the shell.The lock ring was removed from the shell and reinstalled.The liner and the shell were able to be assembled as intended.Device history record (dhr) was reviewed and no discrepancies were found.Root cause remains unchanged.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHELL 50 MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12539620
MDR Text Key273544839
Report Number0002648920-2021-00291
Device Sequence Number1
Product Code KWY
UDI-Device Identifier00889024115422
UDI-Public(01)00889024115422(17)280831(10)64149891
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K833991
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00500105000
Device Lot Number64149891
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/06/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/29/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
CAT# 00500105028 LINER 28 MM LOT# 63900490; CAT# 00500105028 LINER 28 MM LOT# 63900490
-
-