• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ALTRX NEUT 32IDX54OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ALTRX NEUT 32IDX54OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Model Number 1221-32-054
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Insufficient Information (4580); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 09/10/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient operated 2017 (b)(6) clinic.Reoperated due to liner disassociation.Doi: 2017, dor: unknown, unknown side.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthese joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: b3 and h6 (patient).
 
Event or Problem Description
A.Date of event: 2017 c.What is the affected side involved in this event? no information d.Was surgery time extended? if yes, what is the duration of the delay? no delay e.Was the cup available for return? yes f.For reports of disassociation we are required to enter the acetabular cup even if it was not revised.Can you please provide the acetabular cup product / lot information? yes, it has been involved in the first report due to the nature of the reported event, and in accordance with our procedures, we are required to request the following information to assist in the investigation.Please could you let me know if any/none of this information is available: - 1.Radiographs/x-ray films please provide all x-rays relevant to the reported event for example - pre-operative, post primary, pre-revision and intervening time points.Provided in first report 2.Patient demographics the following are the types of patient demographics we require: - - relevant comorbidities, date of implantation, date of revision, operative side, age at the time of the procedure, weight, height, and activity level.2017 3.Implant insertion op notes:- please provide all notes relevant to the reported event for example: - - reason for implant insertion, device labels, surgery report, physiotherapy report, early falls and/or dislocations.No info.4.Implant removal op notes:- please provide all notes relevant to the reported event for example: - - reason for revision, surgeon report and product and lot codes of devices remaining in situ.Liner disassociation, totally loose liner.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary examination of the returned device and provided images found evidence to support disassociation of the liner from the cup while implanted.Depuy considers the investigation closed.Should additional info be received, the investigation will be re-opened as necessary.Device history lot a worldwide complaint database search found no additional related reports against the provided product code/lot number combination.Based on the inability to find any additional related reports against the provided product code/lot number combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.H10 additional narrative:.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRX NEUT 32IDX54OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12539767
Report Number1818910-2021-21227
Device Sequence Number4102866
Product Code LPH
UDI-Device Identifier10603295016120
UDI-Public10603295016120
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K072963
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 09/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1221-32-054
Device Catalogue Number122132054
Device Lot NumberH45118
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2021
Supplement Date Received by Manufacturer09/27/2021
09/23/2021
10/22/2021
12/02/2021
Initial Report FDA Received Date09/28/2021
Supplement Report FDA Received Date10/03/2021
10/05/2021
10/26/2021
12/03/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/15/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX NEUT 32IDX54OD; UNKNOWN HIP ACETABULAR CUP; UNKNOWN HIP FEMORAL HEAD; ALTRX NEUT 32IDX54OD; UNKNOWN HIP ACETABULAR CUP
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-