|
There are three associated devices for this event.They are as below: first device patient identifier: (b)(6).Second device patient identifier: (b)(6).Third device patient identifier: (b)(6).This medwatch is for the first device, patient identifier: (b)(6).The device is not returned, as such a definitive root cause of the reported complaint cannot be determined at this time.Supplemental report(s) will be filed as the information becomes available.
|
|
As reported for this event by the customer, the oleate or serum would not come out through the needle lumen with the needle out.The liquid was coming out with the needle sheathed.This device is used to inject ethanolamine oleate by intra-mucosal bronchial puncture, with the intention of closing small bronchopleural fistulas.It was not possible to use the device.Two other devices from different boxes were tried and had the same issue.There is no reported harm to any patient.
|
|
This supplemental report is being submitted to provide additional information based on the legal manufacturer¿s investigation.The device was returned to olympus for evaluation.The evaluation confirmed that the complete lot number of the device is k9710 (see section d4).During the evaluation, it was confirmed that the needle could extend from the outer tube.It was not possible to inject liquid when the slider was pushed but it was possible to inject liquid when the slider was pulled.The needle tube presented with compressive buckling.No other abnormalities observed that could lead to the reported phenomenon.The legal manufacturer reviewed the device history record (dhr) and no abnormalities detected during the manufacturing of the device.There were no abnormalities detected in the dhr for the following items, which is related to the reported phenomenon: needle extension and retraction, extended length of needle, no bent/kink/crack/scratch on the tube, and injection of the liquid.The device instruction manual contains the following descriptions, and it warns against this reported event (gk6631 rev10): ¿straighten out the instrument before inspecting it.The instrument can be damaged if it is coiled while the handle is operated.Operate the slider slowly, otherwise the tube could buckle.When inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close the biopsy valve, and keep it as straight as possible relative to biopsy valve.Otherwise, the instrument could be damaged.Insert the instrument slowly.Abrupt insertion could damage the endoscope the endoscope and/or instrument.Stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.¿ a review of similar reported complaint in the past, it was likely that the phenomenon occurred due to the compressive bucking on the needle tube.It is unlikely a unit with defective needle is shipped, as nm-401l series undergo 100% inspection for appearance, needle operation and injection.Therefore, the compressive buckling on the needle tube was likely caused when the needle was extended because of the great friction between the outer tube and the needle.It was likely that the friction between the outer tube and the needle increased by the following factors: ·the needle extended/retracted while the tube was coiled in inspection of operation.·the slider was abruptly pushed.·the kink of the tube.·angle of the distal end of the endoscope.
|