| Model Number |
CSE-P-225 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Due to patient confidentiality, the customer was not able to provide patient information or give any details of patients status.Haemonetics field service engineer inspected the device and there were no faults found.On june 11, 2021 haemonetics manufacturing performed a visual inspection of the received bowl from the cell saver® elite set - 125ml and confirmed blood in the inner core with a crack in the inner core base.Although there was no serious injury or harm, past reporting (1219343-2020-00001-01) indicates this particular malfunction on a similar device has been associated with a reported death event.The investigation of 1219343-2020-00001-01 indicated the inner core of the bowl malfunctioned via a crack but did not cause or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing due the event of the past report.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021, haemonetics was notified of air detected early error codes (122, 130, 280, 161, 115) and a required 9 unit transfusion which occurred during a procedure in the (b)(6), utilizing the cell saver® 5 autologous recovery system and cell saver® 5 set - 225ml.
|
| |
|