PHILIPS NORTH AMERICA LLC HEARTSTART XL+ DEFIBRILLATOR/MONITOR Back to Search Results
Model Number 861290
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2021
Event Type  malfunction  
Event Description
It was reported to philips the device did not start and the red "x" is lit for the rfu indicator.There was no patient involvement.
 
Event Description
It was reported to philips the device did not start and the red "x" is lit for the rfu indicator.A philips field service engineer (fse) evaluated the device.The reported issue was confirmed and traced to a faulty inverter pca.The fse replaced the inverter pca.The device passed all performance assurance tests and was placed back into use with the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART XL+ DEFIBRILLATOR/MONITOR
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
elizabeth proctor
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key12540877
MDR Text Key273595892
Report Number3030677-2021-14547
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838023680
UDI-Public00884838023680
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number861290
Device Catalogue Number861290
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/06/2021
Initial Date FDA Received09/28/2021
Supplement Dates Manufacturer Received10/11/2021
Supplement Dates FDA Received11/07/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/19/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-