| Model Number |
0935280 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the bulb syringe was found to have a hair in it.
|
| |
|
Event or Problem Description
|
|
It was reported that the bulb syringe was found to have a hair in it.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed cause unknown.1 sample was confirmed to exhibit the reported failure.The device had not met specifications.Visual evaluation of the returned sample noted one opened (without original packaging), used irrigation bulb syringe.Visual inspection of the sample noted that there was a strand of hair between the barrel and the bulb on the return sample.This does not meet the specification "no hair is permitted on the product.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labelling could not have prevented the reported failure.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was evaluated.
|
| |
|