| Model Number |
CSE-P-225 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Unspecified Blood or Lymphatic problem (4434)
|
| Date of Event |
09/02/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient information is unknown.Haemonetics is reporting based on the amount of blood lost as well as the need for a transfusion.The long empty alarm triggered on the device which stopped the procedure, the device worked as expected.Haemonetics reviewed the data in the history log of the device which showed multiple incorrect operator actions.The disposable sample was discarded, with no sample available for investigation haemonetics is unable to perform an evaluation.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021 haemonetics was notified of a long empty alarm with waste contamination which occurred during a procedure in the united kingdom, utilizing the cell saver® 5 autologous recovery system and cell saver® 5 set - 225ml.The cell salvage blood was reinfused and approximately 600ml was discarded.Donor transfusion was received.
|
| |
|